banner
  • Technical services
  • Turbine

Turbine

Points on working with Turbines

 

    During the intial phases of working with the turbine lubricant spray with high quality must be used. first of all
    shake the oil spray sufficiently and for oil injection use the special nozzle. the oil enters the turbine via the air channel the function continues until clean foam exits from the head of turbine. oil injection should precede Autoc laveing & dental cutling as a repeted operations
    - Before the autoclave, disinfect the turbine surface with spray.
    - Sterilize with an autoclave for 15 minutes at 135 degrees Celsius.
    - Adjust the air inlet to the unit for a turbine mme/sinol of the model BD2 (two holes) 2/2 PSI, and for the turbine mme/sinol of the model BD4 (four holes) and the turbine with an AD coupling load of 2/5 (PSI). Remember that in addition to the water filter, the air purifier and humidifier should be installed inside the unit.
    - Always use a standard bur with a diameter of 1.59- 60 mme/sinol. Even when the turbine is not working, the bur should be inside the turbine.
    - Once the turbine has finished, clean the water channel with air.
    * Tips that can damage the turbine:

    - Damage to the upper part of the turbine.
    - Press the button to remove the bur before complete stop.
    - Use of old and non-standard burs and long burs.
    - Too much lateral pressure to turbine during cutting.
    - Inappropriate air pressure.
    - The entry of humidity into the turbine.
    - Creation of sediment in the cartridge due to the use of low-quality oil spray.
    - Immerse in disinfectants

 


ادامه مطلب
Address: No. 19, first floor, Ehsan complex, north kargar Ave. Tehran, Iran

Tell :+982188632877 

Fax :+982166380186
© Copyright 2017 Aba Pezashk Abzar ,ALL RIGHTS RESRVED